Multifunctional Reference Standard
Accuracy class 0.02, 0.05
National Registry No 52854-13 (extended until 21.11.2027)
Calibration interval 1 year
The Standard of the first grade according to the State Calibration Scheme stated by Rosstandart for the instruments measuring electrical quantities, Order No 1436
Energomonitor-3.1KM can be used independently, together with a PC that provides extended functionality, or as part of a test system. Typical customers are industrial metrological labs and national centers of standardization, metrology and certification.
The closest analogue: RD-30 series of three-phase Electricity Energy Reference Standards produced by Radian (“Radian Research, Inc.”, the USA)
Measured parameters | Measuring range | Intrinsic measurement error | |
Accuracy class 0.02 |
Accuracy class 0.05 |
||
RMS voltage | 0.1 to 960 V (Unom=1,2,5,10,30,60,120,240,480,800 V) | ±0.01% | ±0.02% |
RMS current | 5 mA to 120 A (Inom= 0.05, 0.1, 0.25, 0.5, 1, 2.5, 5, 10, 25, 50, 100 A) | ±0.01% | ±0.02% |
500 mA to 3600 A (Inom = 10, 100, 1000, 300, 3000 A) |
±0.2 ... 2.0% (depending on the type of current sensors) |
||
Phase angle between fundamental harmonics of phase voltages | 0° to 360° | Absolute ±0.01° | |
Phase angle between fundamental harmonics of current and voltage in the same phase | 0° to 360° | ||
Active power | At 1.2∙Inom and 1.2∙Unom | Direct measurements | |
±0.01% | ±0.05% | ||
Measurements via current sensors | |||
±0.1 to 2.0% | ±0.1 to 2.0% | ||
Reactive power | At 1.2∙Inom and 1.2∙Unom | Direct measurements | |
±0.03% | ±0.1% | ||
Measurements via current sensors | |||
±0.2 to 2.0% | ±0.2 to 2.0% | ||
Power factor | -1.0 to +1.0 | Absolute | |
±0.001 | |||
AC frequency | 40 to 70 Hz | Absolute | |
±0.001 Hz | ±0.003 Hz | ||
Negative sequence and zero sequence voltage ratios | 0 to 50 % | Absolute ±0.05 | |
Total Harmonic Distortion of voltage (current) Voltage (current) harmonic of order n (in reference to fundamental KU(n) or KI(n)), where n = 2…50 |
0 to 49.9% | Absolute ±0.003% (KU(n) or KI(n) <1.0 %) |
|
Relative ±0.3% (KU(n) or KI(n)≥ 1.0%) |
|||
Voltage (current) interharmonic of order m KU(m) or KI(m), where m=0.5…50.5 | 0 to 49.9% | Absolute ±0.006% (KU(m) or KI(m) <1.0 %) |
|
Relative ±0.6% (KU(m) or KI(m) ≥ 1.0%) |
|||
Ratio error of instrument voltage and current transformers δ | 0.2 to 20% | Absolute | |
±0.002% | ±0.005% | ||
Phase error of instrument voltage and current transformers Δ | 0.1′ to 180º | ±0.1′ | ±0.2′ |
Flicker short-term severity | 0.25 to 10 | Relative 5.0%(ΔU/U≤20%) |